Alpha-Beta Radiation Detector BDAB for PM1403
Alpha-Beta radiation detector performs the following functions: measurement of α- β- radiation flux density, search for α- β- radiation sources.
BDAB can be used by radiological and isotope tracing laboratories and customs officers, first responders, police, emergency teams.
Operation Principle
A built-in detector based on the proportional counter is used to register a- and β- radiation flux density (DER, count rate). Data communication channel enables information exchange with Spectroscopic Radiation Monitor РM1403 / Display and Control Unit PM1403 by RS-485 interface.
BDAB enables the user to set and constantly monitor the threshold level of α- and β- radiation fluxes through the whole measurement range and record the set threshold level of α- and β- radiation density values in the instrument’s non-volatile memory.
Spectroscopic Radiation Monitor PM1403 Operating Mode
When the instrument is connected to the Spectroscopic Radiation Monitor РМ1403 / Display and Control Unit РМ1403 / Display and Control Unit РМ1403-01 or PC it enables:
α- measurement of β- radiation flux density
α- measurement of β- radiation flux density
search for α- and β- radiation sources
program operation modes
Radiation Detectors Set PM1403
BDG2 – GM detector, gamma radiation detector
BDG3 – Csl(Ti) detector, gamma radiation detector
BDN – He-3 detector, neutron radiation detector
BDAB – proportional counter, alpha-beta radiation detector
Application
Specifications | |
Detector | proportional counter |
Instrument sensitivity: | |
α- radiation (239Pu) | no less than 3.0 impulses∙cm2 |
β- radiation (90Sr+90Y) | no less than 2.0 impulses∙cm2 |
α-radiation flux density measurement range | from 1.0 to 5∙105 min-1cm-2 |
α- radiation flux density accuracy: | |
within range from 1.0 to 5∙105 min-1·cm-2 | ± (20 + А/φ) %, where А – coefficient equal to 10 min-1·cm-2; φ – measured density of α- radiation flux |
β- radiation flux density measurement range | from 10.0 to 106 min-1·cm-2 |
β- radiation flux density accuracy: | |
within range from 10,0 to 106 min-1·cm-2 | ± (20 + А/φ) %, where А – coefficient equal to 100 min-1·cm-2; φ – measured density of α- radiation flux |
Energy range | from 0.15 to 3.5 MeV (β- particles) |
Ingress protection | IP64 |
Data exchange | RS-485 or USB interface |
Power supply | 3.6 (minus 0.6; +0.7) V |
Drop test on concrete floor | 0.7m |
Weight | 0.45 kg |
Dimensions | 71х45х130 mm |
Mean service lifetime | 10 years |
Mean time between failures | 20 000 hours |
Mean recovery time | 60 min |
Operating temperature | from -20 to +50 °С |
Humidity | up to 95 % at 35 °С |
Atmospheric pressure | from 84 to 106.7 kPa |