Gamma Radiation Detection Unit BDG2 for PM1403
Gamma Radiation Detector Detection BDG2 performs the following functions when connected to the Spectroscopic Radiation Monitor РМ1403: indication of the measured value of DER and DE of gamma radiation; indication of average count rate and measured DER value; programming of BDG2 operation modes.
BDG2 can be used by radiological and isotope tracing laboratories, customs officers, emergency services to search and detect ionizing radiation sources.
Operation Principle
Geiger-Muller counter is used to measure DER and DE of photon radiation. Operation algorithm enables continuous measurement process, processing of the results and quick adaptation to radiation intensity changes. BDG2 in the programming mode enables recording data into non-volatile memory and continuous control over two DER and DE threshold values, as well as sound and light alarming when set thresholds are exceeded.
Operating Mode
When the instrument is connected to the Spectroscopic Radiation Monitor РМ1403 / Display and Control Unit РМ1403 / Display and Control Unit РМ1403-01 or PC it enables reading from BDG2:
DER and DE measurement data
DER measurement statistical error
DE accumulation time
BDG2 number
DER and DE threshold levels values
sound alarm state (On/Off)
Recording into BDG2 the following information:
DER and DE threshold levels values
command of accumulated DE and DE
accumulation time resetting
command sound alarm “On/Off”
Radiation Detectors Set PM1403
BDG2 – GM detector, gamma radiation detector
BDG3 – Csl(Ti) detector, gamma radiation detector
BDN – He-3 detector, neutron radiation detector
BDAB – proportional counter, alpha-beta radiation detector
Application
Specifications | |
Detector | Geiger-Muller tube |
Dose equivalent rate (DER) measurement range | from 0.1 μSv/h to 10.0 Sv/h |
DER measurement accuracy | ± (20+K/Ḣ)%, where Ḣ – DER value, μSv/h; К – coefficient = 2,0 μSv/h |
Dose equivalent (DE) measurement range | from 0.1 μSv/h to 9999 mSv |
DE measurement accuracy | ± 10 % |
Thresholds | 0.01 to 9999 mSv (DE thresholds) |
Energy range | from 0.03 to 3.0 MeV |
Energy response relative to 0.662 MeV: | |
within energies range from 30 to 48 keV | minus 40 % |
within energies range from 48 keV to 3.0 MeV | ± 25 % |
Data exchange | USB or RS-485 interface (with BDOI or PC) |
Power supply | 3.6 (minus 0.6; +0.7) V |
Ingress protection | IP65 |
Weight | 0.11 kg |
Dimensions | 162 x 40 mm |
Mean time between failures | 20000 hours |
Mean service lifetime | 10 years |
Mean recovery time | 60 min |
Operating temperature | from -20 to +50 °С |
Humidity | up to 95 % at 35 °С |
Atmospheric pressure | from 84 to 106.7 kPa. |